Imuflex Wb-sp Blood Bag System with Integral Whole Blood Leukocyte Reduction Filter (saving Platelets) with Diversion Blood Sampling Arm

IMUFLEX WB-SP BLOOD BAG SYSTEM WITH INTEGRAL WHOLE BLOOD LEUKOCYTE REDUCTION FILTER (SAVING PLATELETS) WITH DIVERSION BLOOD SAMPLING ARM- dextrose monohydrate, trisodium citrate dihydrate, anhydrous citric acid and sodium phosphate, monobasic, unspecified form solution
Terumo Corporation

NDC 53877-010-01

CPD SOLUTION
For the collection of 500 mL of Whole Blood

FULL PRESCRIBING INFORMATION *Sections or subsections omitted from the Full Prescribing Information are not listed [includes sections 4, 6, 7, 8, 9, 10, 12, 13, 14, 15 and 17].

1. Imuflex Wb-sp Blood Bag System with Integral Whole Blood Leukocyte Reduction Filter (saving Platelets) with Diversion Blood Sampling Arm Indications and Usage

1.1. Read these instructions carefully before use.

1.2. Rx ONLY.

1.3. Intended for the collection, processing and preservation of Whole Blood. Not intended for direct intravenous infusion.

1.4. For the collection of 500 mL ±10% Whole Blood.

1.5. Integral Diversion Blood Sampling Arm is intended to divert and obtain donor samples for laboratory testing prior to collection of the Whole Blood unit.

1.6. Integral filter unit intended for leukocyte reduction of Whole Blood up to 8 hours after blood collection when Whole Blood is stored at ambient temperature.

2. DOSAGE AND ADMINISTRATION

2.1. Prepare the blood bag following your institution’s standard operating procedures.

2.2.1. Materials Needed : Evacuated blood collection tubes (glass or plastic)

2.2. Make a loose knot in the donor tubing below the “Y” and CLIKTIP® (inline closure device) unless alternate methods are used to seal the tubing at the end of collection.

2.3. Temporarily clamp donor tubing between the phlebotomy needle and the “Y”.

Fig. 1
(click image for full-size original)

2.4. Suspend the collection bag as far as possible below the donor’s arm.

2.5. Apply blood pressure cuff or tourniquet to donor’s arm. Disinfect site of phlebotomy. If blood pressure cuff is used, inflate to approximately 60 mmHg.

2.6. Remove the needle cover and perform phlebotomy. Remove the temporary clamp on the donor tubing to permit blood flow into the Diversion Blood Sampling Arm pouch.

2.6.1. CAUTION: Do not touch the needle after removing the needle cover.

2.7. Secure the needle safety device in place following the device instructions provided on the reverse side.

2.8. Secure donor tubing to donor’s arm.

2.9. Position the diversion pouch with the notches up and the Tube Holder/Luer Adapter assembly down. When the level of blood in the pouch is approximately in line with the notches, the diversion pouch is full. (Fig. 2a)

2.9.1. NOTE: The approximate fill volume of the pouch at the notches is 35 mL.

2.10. Permanently seal the tubing between the “Y” and the diversion pouch to maintain a closed system using an aluminum clip or a tube sealer approved for use with tubing connected to a donor (Fig. 2b).

2.10.1. CAUTION: Do not use a dielectric tube sealer to seal the tubing while the needle is connected to the donor’s body unless it is approved for such a purpose.

2.11. To initiate blood flow into the collection bag, break the CLIKTIP between the “Y” and the collection bag.

2.12. To avoid clot formation, collect samples as soon as possible from the diversion pouch as follows (Fig. 2b).

2.12.1. CAUTION: Do not collect donor test samples until the tubing between the “Y” and the diversion pouch is permanently sealed.

2.12.2. Break the CLIKTIP in the tubing below the pouch to open the pathway for sampling. (Fig. 2b)

2.12.3. Position the diversion pouch with the notches up and the Tube Holder/Luer Adapter assembly downward. Assure that any air in the pouch is at the top and will not enter the blood collection tubes.

2.12.4. Open the Tube Holder lid and insert blood collection tube firmly into the tube holder; when full, remove sample tube from holder. Repeat to collect additional samples.

2.12.5. NOTE: The pouch may be removed after the donor test samples are collected. A second seal must be made between the diversion pouch and the permanent seal prior to removing the pouch.

Figs. 2a and 2b
(click image for full-size original)

2.13. Mix blood with anticoagulant in the collection bag and continue to mix at several intervals during collection and immediately after collection. If using an automated mixer, follow manufacturer’s instructions.

2.14. Collect labeled volume of blood 500 mL ±10%.

2.15. When the desired amount of blood has been collected, seal the tubing or tighten the loose knot (white knot) prepared in Step 2.2. Make a second seal between the first seal or knot and the “Y”. Various methods may be used to seal tubing.

2.16. Release pressure on the donor’s arm and remove the needle into the needle safety device following the device instructions provided on the reverse side. Sever the donor tubing between the two seals previously made below the CLIKTIP and “Y”.

2.16.1. CAUTION: Discard the Diversion Blood Sampling Arm and phlebotomy needle/donor tubing according to institutional standard operating procedures.

2.17. Seal and remove donor tubing from collection bag or strip donor tubing as follows:

2.17.1. To obtain a quality control prefiltration sample, strip blood from donor tubing into collection bag, mix well, and allow tubing to refill; repeat once. To prevent the blood from clotting in the tubing, work quickly as possible. Leave an adequate length of tubing containing the well-mixed anticoagulated whole blood attached to the collection bag.

2.17.2. Or, to maximize collection recovery, strip the tubing, mix well and seal tubing close to the collection bag without refilling. To prevent the blood from clotting in the tubing, work quickly as possible. Remove tubing from collection bag.

2.18. The blood filtration is executed according to the following operation. (see Fig.3 and Fig.4)

2.18.1. NOTE: Wait 1 hour after collection before filtering. Filtration can be accomplished when blood is stored at room temperature for up to eight hours after collection.

2.18.2. NOTE: The band guides used to bundle the coiled tubing to prevent entanglement during shipping should be removed prior to filtration (not depicted).

Fig. 3
(click image for full-size original)

Fig. 4
(click image for full-size original)

2.19. Mix the unfiltered whole blood unit by inverting the collection bag #1 (pre-filter bag) several times.

2.20. Hang the collection bag #1, and extend the filtration set to 85 ± 2.5 cm or 33.5 ± 1 inch. Verify that the filter is vertical and ensure all tubing is freely suspended. Position and support bags #2 on a flat surface (Fig. 4).

2.20.1. NOTE: This distance is measured from the outlet of the collection bag to the position of the supported bags as indicated in Fig. 4.

2.20.2. NOTE: Verify tubing has been removed from the tube guides and extends freely.

2.21. Break the CLIKTIP at the outlet of the collection bag #1, and start filtration.

2.21.1. CAUTION: Do not squeeze the collection bag during filtration.

2.21.2. CAUTION: Do not squeeze or apply pressure on the filter while it is attached to the bag containing the filtered blood.

2.21.3. CAUTION: If for any reason whole blood passes through the one-way valve, stop filtration immediately and consider the product as non-leukocyte reduced.

2.22. Filtration ends when the collection bag is empty or blood flow has stopped. Close the WHITE clamp above the filter.

2.23. Expel air through the bypass line and back into the empty collection bag #1 by holding the primary bag #2 upright so that the air will be closest to the ports. Gently squeeze the primary bag containing the leukoreduced whole blood until filtered blood reaches the “Y” below the filter. To ensure ease of air removal ensure primary bag #2 is below the filter outlet.

2.24. Open the WHITE clamp.

2.25. Recovery of filtered Whole Blood is complete when the collection bag and the inlet side of the filter have drained or blood flow has stopped. Close the WHITE clamp.

2.26. Seal the tubing as close as possible to the “Y” below the filter and properly dispose of the filter and collection bag #1.

2.27. Strip the post filter tubing into the primary bag #2, mix well, and allow tubing to refill; repeat once. Make an appropriate number of segments of anticoagulated blood for testing by sealing on or near the X marks. Leave segments attached to the filtered whole blood unit.

2.28. Mix CPD solution and whole blood thoroughly.

2.28.1. NOTE: Whole blood in CPD may be stored for up to 21 days at 1-6°C.

Page 1 of 3 1 2 3

All MedLibrary.org resources are included in as near-original form as possible, meaning that the information from the original provider has been rendered here with only typographical or stylistic modifications and not with any substantive alterations of content, meaning or intent.

This site is provided for educational and informational purposes only, in accordance with our Terms of Use, and is not intended as a substitute for the advice of a medical doctor, nurse, nurse practitioner or other qualified health professional.

Privacy Policy | Copyright © 2023. All Rights Reserved.